Blood on-line detector calibration file

The calibration of blood radioactivity curve measured using on-line detector ("blood pump") requires a file containing the measured calibration coefficients. PET physicists are responsible of the calibration measurements and the updating of the calibration file, currently in /pet-storage/Lab/plasma/bsampler_calibration/pump_cal.dat (UNIX) or S:\Lab\plasma\bsampler_calibration\pump_cal.dat (Windows).

The calibration file is in ASCII text format, and it contains the calibration dates, and the detector-to-gamma counter coefficients for each of the three different on-line detectors (3rd one was installed during summer 2004), and the common gamma counter-to-pet coefficient.

The contents of the on-line detector calibration file on autumn 2004 was:

# YYYY-MM-DD	Pump1(ecat)	Pump2(ge)	Pump3(allogg)	WizardGammacounter
1900-01-01	51		47		1		0.00270
1997-07-01	47		42		1		0.00270
1999-12-15	43		40		1		0.00270
2000-01-13	46.5		42.5		1		0.00270
2000-09-06	45.9		44.6		1		0.00288
2000-11-15	44.9		42.5		1		0.00288
2000-12-05      48.5            45.4            1               0.00288
2001-02-02	46.0		41.6		1		0.00288
2001-03-28	47.8		44.3		1		0.00288
2001-06-07	48.1		43.5		1		0.00288
2001-09-18	42		37		1		0.00288
2001-11-14	44.5		41.4		1		0.00288
2001-12-13	47.5		40		1		0.00277
2002-02-07	47.5		144		1		0.00277
2002-02-26	45.7		133		1		0.00277
2002-03-12	45		40		1		0.00277
2002-12-12	45		41		1		0.00277
2003-06-07	45		42		1		0.00277
2003-08-20	44		41		1		0.00278
2004-01-16	47		41		1		0.00278
2004-09-09	47		41		14		0.00278
2004-09-29	45		40		14		0.00278

For accurate calibration, also the isotope of the tracer must be known to apply a correct branching ratio correction. All calibration steps  can be done using program blo2kbq.