edecay 1.3.4  © 2002-2012 by Turku PET Centre
   
  Reports and changes the correction for physical decay in PET image and
  sinogram data in ECAT 6.3 and 7.x format.
   
  Usage: edecay [Options] <Image or Sinogram>
   
  Options:
   -dd
       Correct PET data for physical decay.
   -rd
       Remove decay correction.
       Make sure that PET is decay corrected before this operation.
   -i=<O-15|N-13|C-11|F-18|Ge-68|Ga-68|Br-76|Rb-82|Cu-62>
       Changes or specifies the isotope.
       Any previous decay correction must be removed before this operation.
       Note that image may need to be recalibrated if isotope is changed;
       at least branching fraction needs to be corrected.
   -dc
       Add decay correction factors to the image data.
       This option does not add decay correction to the pixel values.
   -rc
       Remove decay correction factors from the image data.
       This option does not remove decay correction from the pixel values.
   -h or --help
       Print this message and exit.
   --version or --build
       Print software build information and exit.
   --silent
       Program works silently, printing only error and warning messages.
   --verbose
       Program prints more information about what it is doing.
   
  If no option is specified, program reports the current state of decay
  correction, but makes no changes to the data file.
   
  See also: 
ecattime, 
eframe, 
egetstrt, 
esetstrt, 
ecatunit
   
  Keywords: ECAT, image, physical decay, modelling, simulation
   