Dispersion of the input function
The measured blood curve after a radiotracer bolus is smeared out, because of inhomogeneous velocity fields in the vessels and in the catheter and detector assemblies. Also sticking of radiotracer to the tubing may a add to this dispersion effect, which differs for the measured and true input function to the region of interest.
Dispersion correction with traditional methods adds noise to the blood curve. Regardless of that, dispersion correction should not be accompanied by smoothing (Wollenweber et al., 1997).
Dispersion in the detector system
External dispersion of the input function can be determined by measuring the rising/falling edges of the detector system's response to an input step function (Iida et al., 1986).
Internal dispersion
The internal (physiological) dispersion (caused by human vascular system) between the radial artery and the brain is 4-6 sec (Iida et al., 1986).
Dispersion correction in Turku PET Centre
In Turku PET Centre, when the dispersion correction is necessary, it is usually done automatically by the blood data processing software: first, the external dispersion is corrected, and secondly the internal dispersion is corrected, if an estimate of internal dispersion time constant is available.
In Turku PET Centre, the external dispersion time constant has been estimated to be 2.5 sec for the assemblies that are currently in use. If tubing is changed, the dispersion time constant should be measured again, or asked from the PET physicists.
Internal dispersion time constant of 5 sec has been used for the brain studies.
The corrections are done using
disp4dft.
To continue with the above example, the command could be:
disp4dft off uo0268blo.kbq 2.5 uo0268ab.kbq