PET data

Positron emission tomography (PET) is an imaging method for quantitatively measuring biochemical and physiological processes in vivo by using radiopharmaceuticals and by measuring the annihilation radiation using a coincidence detecting technique.

Physiologically active compounds are tagged with positron-emitting isotopes of oxygen, nitrogen, carbon or fluorine, while maintaining their biological properties. The labeled compounds are administered intravenously in tracer amounts (e.g., in nanomolar or picomolar quantities). Local rates of accumulation, fixation, or turnover of tracer in tissue are measured in absolute units of tissue concentrations from the cross-sectional PET images and represent quantitatively regional tracer tissue concentrations. From the tissue response to an arterial tracer input function, tracer kinetic models are developed that relate the kinetics of a tracer in tissue to the physiological process under study, allow description of the tissue kinetics in mathematic terms, and enable an operational equation to be formulated by which absolute rates of blood flow and metabolism can be calculated.

PET scanner can deliver quantitative images of radioactivity (Bq/voxel, or Bq/mL). The chemical identity of the labeled molecules is only known at the moment of injection. Due to metabolism in blood, in the tissue region of interest, and in other tissues, different labeled compounds are formed in the course of time. The relative amounts of different radioactive compounds in plasma can be measured from arterial blood samples as a function of time.

Quantitative images

PET scanner measures concentration of radioactivity (Bq/mL) as a function of time, producing quantitative 4D images that are stored in PET image files.

Data is quantitative, if signal-to-noise is sufficient, and all the necessary corrections are implemented (attenuation, scatter, dead-time, calibration, movement, partial volume effect, etc).

Countrate and contrast phantoms
Countrate and contrast phantoms.

Image visualization

The PET images that you see in publications and presentations, are for visualization (or entertainment) purposes only. Color scales have been enhanced at the expense of quantitation, just to please human eye and to emphasize certain structures and differences. Even if image is not manipulated on purpose, the process of producing image for visualization destroys the precision in the original PET image data. Concentrations in PET image file are real numbers in real units, but we have to assign a range of concentration values to a certain color to show a color image. Different color scales (gray, rainbow, hot metal, …) can be selected, and highest or lowest pixel values can be cut off.

image visualization

Concentrations in PET

PET measures the concentration of radioactive label, for example isotope 18F. However, we are interested in concentration of a ligand, carrying the label, in tissues (pmol/mL). Measurement is based on observing the emitted gamma rays as the label decays. Ligand itself is not decaying. Measured radioactivity concentration must be corrected for the physical decay — only after that the radioactivity concentration is in proportion to ligand mass.

Radioactivity concentration at t=0, C(0), can be calculated from equation 1, where λ is:

As the label is decaying, the radioactivity measurement becomes more difficult. To compensate that, the duration of PET “frames” are extended to collect more data for each concentration measurement. Halflife of the isotope label, physiological clearance of the ligand, and sensitivity of the scanner determine how long the ligand concentration can be measured with PET.

Radioactivity is proportional to mass

After decay correction, radioactivity concentration is proportional to the ligand concentration. The proportionality constant (specific radioactivity) is measured at the time of tracer administration (t=0). With that we can convert the measured radioactivity concentration (Bq/mL) to the chemical concentration of the ligand in picomoles per litre.

radioactivity versus mass

Tissue uptake

The radioligand concentration in tissue is dependent on

radioactivity versus mass

Ligand concentration in tissue is related to the things which we are interested in studying, such as perfusion, or metabolic rate of glucose, or receptor density, etc. These are properties of the ligand as well — how specific the radioligand uptake is to the phenomenon that we want to measure. For example, in skeletal muscle, low perfusion is limiting the uptake of most radioligands, even though our intension is to measure something else than perfusion, like protein synthesis. In the brain, multidrug efflux transporters may affect ligand brain uptake (Kroll et al., 2014).

Radioligand is usually also a target of metabolism, at least in liver, and if radioactive metabolites are released to circulation, those may obscure the results if taken into the region of interest.


See also:



References:

Bendriem B, Townsend DW (Eds.): The Theory and Practice of 3D PET. Springer, 1998, 168 p, ISBN 978-90-481-5040-3.

Khalil MM (ed.): Basic Science of PET Imaging. Springer, 2017. DOI 10.1007/978-3-319-40070-9.

Teuho J, Johansson J, Linden J, Hansen AE, Holm S, Keller SH, Delso G, Veit-Haibach P, Magota K, Saunavaara V, Tolvanen T, Teräs M, Iida H. Effect of attenuation correction on regional quantification between PET/MR and PET/CT: a multicenter study using a 3-dimensional brain phantom. J Nucl Med. 2016; 57: 818-824.



Tags: , ,


Created at: 2013-12-17
Updated at: 2017-05-17
Written by: Vesa Oikonen, Tuula Tolvanen