Delay time correction

What is time delay?

Although the shape of radioactivity concentration curve is similar in all arteries, the measured radioactivity peak seems to arrive at different times to the tissue (PET) and to the blood samples, because of sample tubing and different distance and blood velocity in different arteries. In many PET analyses, the effect of this difference in tracer arrival times is insignificant, but in some studies, particularly all that are done with radiowater or oxygen, or where a full compartmental model is fitted, the delay time must be considered.

How to correct for time delay?

What data do we need?

For delay correction, corrected blood curve measured using on-line detector and PET countrate curve (*.cr) are normally used.

It may be possible to use manually sampled blood or plasma curve instead of on-line detector data, if manual samples have been taken at 10-15 s intervals (at max), and the initial increasing phase of the curve contains several samples.

Also, the countrate curve can be replaced with regional TACs (see below) or "head curve", an average TAC from the dynamic image or sinogram made with ecathead, if PET time frames are short enough (10-15 s or less) in the beginning of the PET scan.

Countrate curve must not be used in certain cases

If the heart, lungs or aorta is inside the PET image volume (or even close to it in 3D studies, in which the spill-over artefacts may be significant), then the use of countrate or head curve is discouraged: In the heart and large arteries the initial radioactivity concentration is relatively high and appears considerably sooner than in the tissue of interest. This may cause a bias of a few seconds in the estimated delay time. For example in perfusion studies of renal cortex and tumours in oral cavity, a region-of-interest TAC should be used in the delay time correction in place of count-rate curve.

Software

In most cases, the time delay correction is included in the blood data pre-processing scripts for oxygen-15 labelled or other radiotracers. Alternatively, certain model analysis software makes automatically or optionally also the time delay correction, for example fit_h2o.

If you need to do the time delay correction by yourself, you can use fitdelay. However, different tracers may require specific settings for this program to work reliably.

In all cases, you must always visually check that time delay correction was successful!

If time delay is known from elsewhere, e.g. by visual inspection, the time delay can be corrected using dfttime.

If it does not work

The most common reasons for failure in automatic delay correction are:

Note that the failure is not always obvious, but is only seen as biased results, unless you check the correction by plotting fitted input curve and tissue curve together.

Do it only after metabolite correction

The fractions of authentic tracer and metabolites are measured at sample times that are relative to the sample times of the total plasma or blood curve. Therefore, the metabolite correction must be made before the sample times of plasma or blood curve are changed.

If you use fitdelay to make the delay correction, you can correct all plasma and blood curves belonging to the same study by entering the filenames to the end of the command line.



References:

Iida H, Higano S, Tomura N, Shishido F, Kanno I, Miura S, Murakami M, Takahashi K, Sasaki H, Uemura K. Evaluation of regional difference of tracer appearance time in cerebral tissues using [15O]water and dynamic positron emission tomography. J. Cereb. Blood Flow Metabol. 1988; 8: 285-288.

Iida H, Kanno I, Miura S, Murakami M, Takahashi K, Uemura K. Error analysis of a quantitative cerebral blood flow measurement using H215O autoradiography and positron emission tomography, with respect to the dispersion of the input function. J. Cereb. Blood FlowMetab. 1986; 6: 536-545.

Meyer. Simultaneous correction for tracer arrival delay and dispersion in CBF measurements by the H215O autoradiographic method and dynamic PET. J Nucl Med 1989; 30:1069-1078.

van den Hoff et al. Accurate local blood flow measurements with dynamic PET: fast determination of input function delay and dispersion by multilinear minimization. J Nucl Med 1993; 34:1770-1777.



Valid XHTML 1.0! Valid CSS!